Low Differential Pressure Calibration: Challenges and Best Practice

Home › Resources › Low Differential Pressure Calibration

Low differential pressure calibration covers ranges from a few pascals up to about 10 kPa, used by cleanroom monitors, HVAC sensors, filter gauges and DP transmitters. At these levels, temperature changes, tiny leaks and air movement can create errors larger than the instrument's tolerance, so stable pressure control and careful set-up matter more than at higher pressures.

By TIS Instruments engineering team · Last updated 26 November 2026

Where Low DP Instruments Are Used

  • Cleanroom room-pressure monitors (pharmaceutical, semiconductor, hospitals)
  • HVAC duct, fan and filter differential pressure
  • Fume hoods, isolators and biosafety cabinets
  • Data centre containment and airflow
  • Draft gauges and low-range DP transmitters

Five Common Problems

  1. Temperature changes – warm hands on a hose, sunlight or an air-conditioning draught change the air temperature in the line, and the pressure moves with it.
  2. Leaks – a leak that is invisible at 10 bar makes a 50 Pa reading drift continuously.
  3. Volume and settling – long or wide hoses add volume, so each point takes longer to settle.
  4. Resolution – the reference must resolve fractions of a pascal to calibrate a ±25 Pa instrument properly.
  5. Manual control – adjusting a low pressure calibration pump by hand often overshoots, so the technician has to start again.

Best Practice

  • Let instruments and hoses reach room temperature; avoid handling hoses during readings.
  • Use short, small-bore, low-permeability tubing and as few joints as possible.
  • Leak-check at the top of the range before you start.
  • Zero both the reference and the device under test with both ports vented together.
  • Shield the set-up from air-conditioning draughts and direct sunlight.
  • Allow each point to settle; record only when readings are stable.
  • Use a reference with enough resolution and accuracy for the range (4:1 rule).

Manual Pump vs Automatic Calibrator

Manual low pressure calibration pump Automatic low pressure calibrator
Setting a point Hand adjustment, then wait Enter the set point; it is controlled
Drift correction Technician re-adjusts Controller keeps correcting
Time per instrument Long Short and repeatable
Electrical reading Separate meter Built in (mA, V, switch)

The TIS APC150-LP automatic low pressure calibrator covers ±0.1 kPa to ±10 kPa at 0.025% or 0.01% FS, with 0.005% FS pressure stability and built-in mA/V inputs. For portable checks, the Precision Low Differential Pressure Calibrator covers ±100 Pa to ±10,000 Pa in a 1.3 kg handheld.

Frequently Asked Questions

What counts as low differential pressure?

There is no single definition, but ranges below about 10 kPa (100 mbar or 40 inH₂O) are usually called low pressure, and cleanroom and HVAC instruments often work between ±25 Pa and ±2.5 kPa.

Why does my low pressure reading keep drifting?

Most drift comes from temperature changes in the test line or small leaks. Let the set-up stabilise, avoid touching hoses, shorten the tubing and leak-check every joint.

Which units are used for low pressure?

Pascal (Pa), kilopascal (kPa), millibar (mbar), inches of water (inH₂O) and millimetres of water (mmH₂O) are common. 1 inH₂O is about 249 Pa. Use our pressure units guide for conversions.

Can I calibrate a DP transmitter’s 4–20 mA output at the same time?

Yes, with a calibrator that reads mA and supplies 24 V loop power, such as the APC150-LP. You apply each pressure point and read the transmitter output in the same step.