Automatic Pressure Calibrator
An automatic pressure calibrator (also called an automated pressure calibrator) generates and controls pressure to each set point by itself, so the technician only enters the test points and records the results. TIS automatic pressure calibrators cover pneumatic, hydraulic and low differential ranges at 0.025% or 0.01% FS, and are designed and built in Singapore.
Showing all 6 results
-

Automatic Hydraulic Pressure Calibrator – AHCP 700
-

Automatic Hydraulic Pressure Calibrator AHCP700-DK
-

Automatic Low Pressure Calibrator – APC150-LP
-

Automatic Pneumatic Pressure Calibrator APC-300
-

Portable Automatic Pneumatic Pressure Calibrator – APC 150 S
-

Portable Automatic Pneumatic Pressure Calibrator APC-150
What Is an Automated Pressure Calibrator?
An automated pressure calibrator is a calibrator with a built-in pressure controller. Instead of turning a hand pump and a fine adjuster until the reading is close, you type the target pressure – for example 25%, 50%, 75% and 100% of range – and the instrument raises or lowers the pressure and holds it steady. While the pressure is stable, it also reads the device under test, such as a transmitter’s 4–20 mA output, and shows the error.
“Automatic” and “automated” pressure calibrator mean the same thing. Some suppliers also call it an automatic pressure controller/calibrator.
How an Automatic Pressure Calibrator Works
Built-in Pressure Generation and Control
The calibrator has a pressure source (an internal pump, or a gas or hydraulic supply), control valves and one or more precision reference sensors. A control loop opens and closes the valves until the reference sensor reads the set point.
Set Point, Stability and Test Sequence
When the set point is reached, the controller holds it within a tight band. On the TIS APC-150 the control stability is 0.002% FS in less than 15 seconds; on the AHCP700-DK it is 0.01% FS in less than 15 seconds with water. With PC software, the full sequence of up and down test points can run without the operator touching a valve.
Benefits Over Manual Calibration
- Faster: no pumping and no fine adjusting at every point, so each instrument takes less time.
- Repeatable: the same set points and the same stability every time, whoever is operating.
- Fewer operator errors: no overshoot, no misreading of a dial.
- Built-in electrical measurement: reads mA (loop powered) and volts, and tests pressure switches.
- Better records: results can be saved and transferred by USB or RS232 for certificates and audits.
TIS Automatic Pressure Calibrator Models
| Model | Medium | Pressure range | Accuracy | Key features |
|---|---|---|---|---|
| AHCP700 | Hydraulic | 0–100, 200, 400, 700 or 1,000 bar (up to 2 internal sensors) | 0.025% or 0.01% FS | 4.3″ colour touch screen, mA/V input, switch test, USB, Windows software |
| AHCP700-DK | Hydraulic (water or oil) | 0–700 bar (70 MPa / 10,000 psi) | 0.025% FS (0.01% optional) | 7″ touch screen, 0.01% FS stability in under 15 s, bench unit |
| APC-300 | Pneumatic | ±2.5 kPa to -90~2,000 kPa (up to 3 sensors) | 0.05%, 0.025% or 0.01% FS | 0.002% FS stability, mA/V input, RS232 |
| APC-150 | Pneumatic | -atm to 300 psi | 0.025% FS (0.01% optional) | Built-in electric pump, 8-hour rechargeable battery, portable |
| APC150-LP | Low differential | ±0.1 kPa to ±10 kPa | 0.025% or 0.01% FS | 0.005% FS stability, for cleanroom and HVAC |
Pneumatic vs Hydraulic Automatic Calibrators
Pneumatic automatic calibrators (APC series) use air or nitrogen. They are clean, quick and ideal for low and medium ranges – from a few pascals up to about 20 bar. Choose them for DP transmitters, pneumatic gauges and switches.
Hydraulic automatic calibrators (AHCP700 series) use oil or water. Liquid barely compresses, so high pressures up to 1,000 bar are reached safely and held steadily. Choose them for high-pressure transmitters, hydraulic gauges and wellhead or process instruments.
For very low ranges, such as cleanroom room pressure or filter DP, see the APC150-LP automatic low pressure calibrator.
Applications
- Calibration laboratories and instrument workshops
- Pressure transmitter and gauge production lines
- Oil and gas, petrochemical and marine instrument maintenance
- Pharmaceutical and semiconductor plants (cleanroom DP)
- Power plants and utilities
Frequently Asked Questions
Is there a difference between an automatic and an automated pressure calibrator?
No. Both terms describe a calibrator with a built-in pressure controller that sets and holds each test pressure by itself. Some makers say “automated pressure calibrator”, others say “automatic pressure calibrator” or “pressure controller/calibrator”. The working principle and the benefits are the same.
Can an automatic pressure calibrator measure 4–20 mA transmitter output?
Yes. TIS automatic pressure calibrators read 0–24 mA (loop powered) and 0–30 V DC, so you can calibrate a pressure transmitter in one step. They also detect the change of state of a pressure switch, normally open or normally closed, without a separate multimeter.
What accuracy do TIS automatic pressure calibrators offer?
Most models are available at 0.025% FS, with 0.01% FS as an option. The APC-300 can also be supplied at 0.05% FS. Pressure stability ranges from 0.002% FS to 0.005% FS depending on the model, which gives steady readings at each test point.
Does an automatic pressure calibrator need an external gas supply?
It depends on the model. The portable APC-150 has a built-in electric pump and generates its own pressure. TIS also offers a pressure controller version of the APC-300 that runs from an external nitrogen (N₂) supply. Tell us your range and site setup and we will recommend the right option.
Is an automatic pressure calibrator portable?
Some are. The APC-150 series is a portable automatic calibrator with an 8-hour rechargeable battery for site work. Hydraulic models such as the AHCP700 and AHCP700-DK are bench units for a lab or workshop, where high pressure and long test runs are needed.
Related Guides
Request a Quote
Not sure which automatic pressure calibrator fits your instruments? Send us your pressure range, medium and accuracy, or call +65 6779 9272. See the full range of pressure calibration equipment or compare with our electric pump pressure calibrators.