Pressure Accuracy Explained: % FS, % of Reading and the 4:1 Rule

Home › Resources › Pressure Accuracy Explained

Pressure instrument accuracy is usually given either as a percentage of full scale (% FS) or as a percentage of reading. A % FS specification gives the same maximum error everywhere in the range, while a % of reading specification gets smaller as the pressure gets lower. To calibrate an instrument reliably, the reference should be at least four times more accurate than the instrument under test.

By TIS Instruments engineering team · Last updated 26 November 2026

% of Full Scale (% FS)

The error limit is a percentage of the instrument’s full range, and it is the same in pounds, bar or pascals at every point.

Example: a 0–100 bar calibrator with ±0.025% FS has a maximum error of ±0.025 bar – at 100 bar and at 10 bar.

At 10 bar, that same ±0.025 bar is 0.25% of the reading. This is why a high-range instrument is a poor choice for low-pressure work.

% of Reading

The error limit is a percentage of the value being measured.

Example: the TIS Baro-Cal barometer calibrator is specified at 0.01% of reading. At 100 kPa the limit is ±0.01 kPa (10 Pa); at 90 kPa it is ±9 Pa.

Comparing the Two

Pressure applied 0.025% FS (100 bar range) 0.05% of reading
100 bar ±0.025 bar ±0.050 bar
50 bar ±0.025 bar ±0.025 bar
10 bar ±0.025 bar ±0.005 bar

Neither is “better” on its own – compare the actual error at the pressures you use most.

Accuracy vs Uncertainty vs Resolution

  • Accuracy – how close a reading is to the true value, within the manufacturer’s specification.
  • Measurement uncertainty – the range within which the true value is expected to lie, stated on a calibration certificate with a confidence level (usually about 95%).
  • Resolution – the smallest change the display can show. High resolution does not guarantee high accuracy.

The 4:1 Rule (Test Uncertainty Ratio)

A widely used rule is that the reference should be at least four times more accurate than the device under test. This keeps the reference’s own error small compared with the tolerance you are checking.

Instrument under test Tolerance Reference needed (4:1)
Process gauge ±1.0% FS ±0.25% FS or better
Test gauge ±0.25% FS ±0.05% FS or better
Transmitter ±0.1% FS ±0.025% FS or better
High-accuracy transmitter ±0.04% FS ±0.01% FS or better

Always compare like with like: convert both specifications to the same units at the same pressure.

Choosing the Right Range

Pick a reference range close to the instrument range. A 0–20 bar reference at 0.025% FS (±0.005 bar) is far better for a 0–10 bar gauge than a 0–700 bar reference at the same % FS (±0.175 bar). TIS calibrators offer multiple ranges and accuracy grades from 0.05% to 0.01% FS – see our pressure calibrators.

Frequently Asked Questions

What does 0.025% FS mean?

It means the maximum error is 0.025% of the instrument’s full-scale range. On a 0–10 bar range, that is ±0.0025 bar at any point in the range.

Is 0.01% FS better than 0.025% FS?

On the same range, yes – 0.01% FS allows an error two and a half times smaller. Across different ranges, compare the actual error in pressure units at the points you use.

What if I cannot reach a 4:1 ratio?

Some high-accuracy instruments cannot be checked at 4:1. In that case, use a reference with the smallest uncertainty available, account for that uncertainty in the pass/fail decision, and document the approach in your procedure.

Does temperature affect accuracy?

Yes. Specifications usually apply within a stated temperature range (for example 0–50 °C). Outside that range, extra temperature error can apply, so check the datasheet.