How Often Should Pressure Instruments Be Calibrated?

Home › Resources › How Often Should Pressure Instruments Be Calibrated?

Most pressure gauges, transmitters and pressure calibrators are calibrated once every 12 months. That is a starting point, not a rule: the right interval depends on how critical the instrument is, how heavily it is used, the conditions it works in, and how much it drifted at previous calibrations. Your quality system should define the interval and the reasons for it.

By TIS Instruments engineering team · Last updated 26 November 2026

What the Standards Say

  • ISO 9001 (clause 7.1.5) requires measuring equipment to be calibrated or verified at specified intervals, or before use, against standards traceable to international or national standards. It does not set a fixed interval.
  • ISO/IEC 17025 laboratories do not set your interval; their certificate records the result on the day of calibration.
  • ILAC-G24 / OIML D 10 gives guidance on methods for setting and reviewing calibration intervals.

Typical Starting Intervals

Instrument Typical starting interval
Process pressure gauges 12 months
Pressure transmitters 12 months (some critical loops 6 months)
Pressure switches (safety-related) As required by the safety system, often 6–12 months
Digital test gauges / master gauges 12 months
Pressure calibrators and controllers 12 months
Deadweight testers 12–24 months, depending on use and history

When to Shorten the Interval

  • The instrument was found out of tolerance at its last calibration.
  • It is used for safety, custody transfer or product quality decisions.
  • It works with vibration, pulsation, high temperature or corrosive media.
  • It has been dropped, over-pressurised or repaired.
  • It is used very frequently as a reference.

When You Can Extend It

  • Several consecutive calibrations show small, stable drift well inside tolerance.
  • The instrument is lightly used in a clean, stable environment.
  • Intermediate checks between calibrations confirm it is still accurate.

Change intervals in small steps and record the evidence.

Simple Interval Review Method

  1. Record “as found” results at every calibration.
  2. If an instrument is found out of tolerance, shorten its interval (for example by 25–50%) and check the impact on products or processes measured since the last calibration.
  3. If it is found well within tolerance several times in a row, consider extending the interval moderately.
  4. Review intervals at least once a year.

Frequently Asked Questions

Is annual calibration mandatory?

No standard makes annual calibration mandatory for all instruments. Twelve months is a common, practical default. Customer contracts, industry regulations or your own risk assessment may require a different interval.

Does a calibration certificate expire?

A certificate does not expire; it shows the instrument’s condition on the calibration date. Your quality system decides when the next calibration is due, and many companies label instruments with a due date.

Should a new instrument be calibrated before use?

Yes, if it will be used for measurements that matter. Many manufacturers supply a factory certificate; for audits, an accredited certificate such as SAC-SINGLAS gives stronger evidence of traceability.

Where can I get pressure instruments calibrated in Singapore?

TIS Instruments provides SINGLAS calibration for pressure instruments from vacuum up to 5,000 bar, including deadweight testers, under accreditation LA-2019-0722-C.

Talk to TIS Instruments

Looking for the right equipment? See our SINGLAS calibration or ask our engineers.