Walk into almost any established calibration laboratory and you will find a deadweight tester on a solid bench: a polished piston, a stack of stainless steel weights, and a technician carefully spinning the weights before taking a reading. Walk into a modern instrument workshop and you are just as likely to see a digital calibrator running an automatic test sequence. Both measure pressure. They just do it in very different ways.
A deadweight tester creates a known pressure by placing calibrated weights on a piston of known area – pressure is simply force divided by area. A digital pressure calibrator measures pressure electronically with a precision sensor. Deadweight testers are a classic laboratory reference with excellent long-term stability; digital calibrators are faster, more portable and easier to automate. Here is how to decide which belongs in your lab or workshop.
Key takeaways
- Deadweight testers generate pressure from mass, area and gravity – a fundamental method.
- Digital calibrators measure pressure with a sensor and are faster and portable.
- Deadweight testers need corrections for local gravity, temperature and air buoyancy.
- Many labs use a deadweight tester as the reference and digital calibrators for daily work.
- Both need regular calibration – deadweight testers are often checked by cross-floating.
In this article
- How a Deadweight Tester Works
- How a Digital Pressure Calibrator Works
- Side-by-Side Comparison
- When a Deadweight Tester Makes Sense
- When a Digital Calibrator Makes Sense
- Using Both Together
- Calibrating Deadweight Testers
- Speed in Practice
- Questions to Ask Before You Decide
- Conclusion
- Frequently Asked Questions
How a Deadweight Tester Works
A deadweight tester (also called a piston gauge) has a precisely made piston inside a closely fitting cylinder. Calibrated weights are placed on the piston. When the system pressure lifts the piston so it floats freely, the pressure underneath equals the force of the weights divided by the piston's effective area.
Because the force comes from mass and gravity, the result must be corrected for:
- Local gravity – gravity varies with location, so the same weights produce slightly different force in different places.
- Temperature – the piston area changes slightly with temperature.
- Air buoyancy – the air around the weights pushes up on them a little.
Done properly, the result is a very stable, well-understood reference pressure.

How a Digital Pressure Calibrator Works
A digital calibrator uses a pressure sensor – often a precision piezoresistive or resonant sensor – whose output is compensated for temperature and linearity, then displayed in your chosen unit. Many digital calibrators also generate pressure and measure mA and volt signals, so they can calibrate transmitters and switches in one step.
Side-by-Side Comparison
| Deadweight tester | Digital pressure calibrator | |
|---|---|---|
| Principle | Force ÷ area (fundamental method) | Electronic sensor |
| Long-term stability | Excellent | Good; needs regular calibration |
| Speed | Slower – weights changed per point | Fast; can be automated |
| Portability | Heavy, needs a level bench | Portable models available |
| Corrections | Gravity, temperature, buoyancy | Built in |
| Operator skill | Higher | Lower |
| Typical use | Reference standard in labs | Workshop, field and lab work |
When a Deadweight Tester Makes Sense
- You run a calibration laboratory and need a very stable, traceable reference.
- You calibrate other high-accuracy instruments, such as digital calibrators and test gauges.
- Your work happens on a bench, and speed is less important than stability.
Pro tip
If you use a deadweight tester, record the local gravity value used in your corrections and keep the weights clean and handled with gloves. Fingerprints and dust add mass and error over time.
When a Digital Calibrator Makes Sense
- You calibrate many instruments and want speed and automatic records.
- You need to work on site, where a deadweight tester is impractical.
- You also need electrical measurements for transmitters and switches.
TIS digital calibrators offer accuracy up to 0.01% FS, from handheld units to automatic hydraulic systems up to 1,000 bar. See our pressure calibration equipment.
Using Both Together
In practice, many laboratories use both. The deadweight tester is the stable reference, calibrated by an accredited lab at long intervals. Digital calibrators do the everyday work and are periodically checked against the deadweight tester. This combines long-term stability with day-to-day speed.
Calibrating Deadweight Testers
Deadweight testers are themselves calibrated, often by cross-floating – connecting the tester to a reference piston gauge, balancing both at the same pressure, and determining the effective area of the tester from the reference.
TIS performs SAC-SINGLAS accredited cross verification of pneumatic deadweight testers from 1.44 to 2,030 psi and hydraulic deadweight testers from 5 to 4,136 bar. Details are on our SINGLAS calibration service page.
Good to know
Not sure what uncertainty you actually need? Our article on pressure accuracy, % FS and the 4:1 rule helps you work out the reference accuracy your instruments require.
Speed in Practice
Consider calibrating a 0–100 bar transmitter at five points up and five points down. With a deadweight tester, each point means selecting and loading the right weights, spinning them, waiting for the piston to float and settle, and then reading the transmitter. With an automatic digital calibrator, the technician enters the points once and the controller works through them, reading the transmitter's mA output at each step.
For a single instrument the difference is a few minutes. For a workshop calibrating dozens of transmitters a week, it becomes hours – which is why most production workshops use digital calibrators for daily work.

Questions to Ask Before You Decide
- What uncertainty do my instruments really need? If your best instruments are 0.1% FS, a 0.025% FS digital reference may be all you need.
- Where will the work happen? Field work rules out deadweight testers.
- How many instruments per week? High volume favours automation.
- Who will operate it? Deadweight testers need more training to use correctly.
- How will I keep it traceable? Plan accredited calibration for whichever reference you choose.
Conclusion
A deadweight tester and a digital calibrator are not rivals so much as partners. If you need a rock-solid laboratory reference, the deadweight tester is hard to beat. If you need speed, portability and automation, a digital calibrator is the practical choice. Many of the best-run labs use both, each for what it does best.
Frequently Asked Questions
Is a deadweight tester more accurate than a digital calibrator?
High-grade deadweight testers can achieve very low uncertainty and are often used as laboratory references. Good digital calibrators come close and are far quicker to use. Compare the stated uncertainty of the specific instruments.
Why do deadweight testers need gravity correction?
The force from the weights depends on local gravity, which varies with location. Using the local value avoids an error of up to a few hundredths of a percent.
What is cross-floating?
Cross-floating connects two piston gauges and balances them at the same pressure, so the effective area of one can be determined from the other.
How often should a deadweight tester be calibrated?
Typical intervals are 12–24 months, depending on use, history and your quality system.
Talk to TIS Instruments
Need the right equipment or an accredited calibration? See our SINGLAS calibration service or ask our engineers.