By TIS Instruments engineering team  ·  27 November 2026  ·  5 min read

A pharmaceutical plant needs to check fifty RTDs on a production line without bringing liquid into a clean area. A calibration laboratory needs to certify short sanitary probes with the lowest possible uncertainty. Both need a temperature calibrator – but not the same one.

A dry block calibrator heats or cools a metal block with drilled wells, so it is portable, clean and quick. A liquid bath circulates water or oil around the probe, giving better stability and uniformity and accepting probes of almost any shape. In most cases the choice is simple: dry block for field work and routine sensor checks, bath for the best accuracy and for short, odd-shaped or multiple probes in a lab. This article explains the trade-offs in detail.

Key takeaways

  • Dry blocks: portable, clean, fast – ideal for field checks and routine work.
  • Liquid baths: best stability and uniformity – ideal for labs and difficult probes.
  • Immersion depth and insert fit matter more in a dry block than in a bath.
  • An external reference probe improves dry block accuracy.
  • Furnaces extend calibration to very high temperatures.

How a Dry Block Calibrator Works

A dry block contains a metal block (often with removable inserts) heated or cooled to a set temperature. You insert the sensor under test into a well that fits it closely. Heat transfers through metal contact, and when the block is stable, you compare the sensor's reading with the calibrator's reference.

Advanced models, such as the TIS HTC and LTC multifunction dry blocks, also include a reference thermometer input and a signal calibrator that reads RTDs, thermocouples, thermal switches and mA signals, with HART communication and data logging.

How a Liquid Bath Works

A bath holds a stirred liquid – water, silicone oil or another fluid – at a controlled temperature. The probe is simply immersed in the liquid. Because the liquid surrounds the probe and is constantly stirred, temperature is very even, and probes of any shape can be calibrated.

TIS ATCS1200 bath and furnace temperature calibration system
The TIS ATCS1200 system combines water and silicone oil baths with a furnace, covering 5 °C to 1,200 °C.

Side-by-Side Comparison

Dry block calibrator Liquid bath
Portability High – carry it to site Low – laboratory equipment
Cleanliness No liquid Water or oil on probes
Stability & uniformity Good Excellent
Probe fit Needs matching well or insert Any shape and size
Short probes Harder (immersion depth) Easier
Speed Fast heating and cooling Slower

When a Dry Block Is the Better Choice

  • Field and on-line checks of RTDs, thermocouples and thermometers.
  • Routine calibration where moderate uncertainty is acceptable.
  • Clean areas where liquids are not allowed.

TIS compact dry blocks include the TC15M (-40 to 150 °C, ±0.3 °C accuracy) and the TC65M (30 to 650 °C, ±0.5 °C accuracy).

When a Bath Is the Better Choice

  • Lowest uncertainty and best uniformity.
  • Short, bent or sanitary probes that do not fit a well.
  • Many probes at once, for example batch calibration in a lab.

The TIS ATCS1200 combines a water bath (5–90 °C) and a silicone oil bath (ambient–220 °C) with ±0.01 °C stability, plus a furnace for 200–1,200 °C. See all temperature calibrators.

Getting Good Results in a Dry Block

Most dry block errors come from heat flowing along the probe instead of into it. To avoid them:

  1. Use the right insert – the hole should fit the probe closely, without forcing.
  2. Immerse deeply – a common guide is at least 15 times the probe diameter.
  3. Use an external reference probe in the same insert for the best accuracy.
  4. Wait for stability at each point before recording.

Pro tip

Keep inserts clean and dry, and never force a probe into a tight hole. A small air gap of just a fraction of a millimetre can add a noticeable error at high temperatures.

Step-by-Step: Calibrating an RTD in a Dry Block

  1. Choose the insert with a hole that matches the RTD diameter, and a second hole for the reference probe if you use one.
  2. Insert both probes to the bottom of the well.
  3. Connect the RTD to the calibrator's input or to a separate readout.
  4. Set the first temperature – for example 0 °C or the lowest point of the sensor's working range.
  5. Wait for stability – the calibrator's stability indicator, or a steady reading for a few minutes.
  6. Record the reference temperature and the RTD reading.
  7. Repeat at the other points, such as 50, 100 and 150 °C.
  8. Calculate the errors and compare them with the sensor's tolerance class.
TIS TC65M compact dry block temperature calibrator
The compact TIS TC65M dry block covers 30 to 650 °C with ±0.5 °C accuracy and a 26 × 155 mm well.

Choosing Inserts for Your Probes

Inserts are the part of a dry block that most often gets overlooked. A well-chosen set covers the probe diameters you use most – for example 3, 4, 6 and 8 mm – plus a hole for the reference probe. Multi-hole inserts let you calibrate several probes together, but every probe must still fit its hole closely.

What About Very High Temperatures?

Above the range of oil baths, furnaces take over. They are used for thermocouples at several hundred degrees and above. The ATCS1200's furnace section, for example, covers 200 to 1,200 °C with ±0.5 °C stability.

Good to know

Temperature calibration at TIS is traceable to national and international standards. TIS's SINGLAS accreditation (LA-2019-0722-C) covers pressure, so temperature instruments receive a traceable calibration certificate. For accredited pressure work, see our SINGLAS calibration service.

Conclusion

If you need to take the calibrator to the sensor, choose a dry block. If you need the best possible numbers, or your probes are short, bent or numerous, choose a bath. Many organisations end up with both: dry blocks for the technicians in the field, and a bath or bath-and-furnace system in the lab.

Frequently Asked Questions

Which is more accurate, a dry block or a bath?

Baths usually offer better stability and uniformity, so they achieve lower uncertainty. Good dry blocks with an external reference probe are accurate enough for most industrial sensors.

Can a dry block calibrate a thermocouple?

Yes. Insert the thermocouple in a well of the right size and read its output with a signal calibrator or the dry block's built-in input.

Why does immersion depth matter?

Heat flows along the probe stem. If the probe is not deep enough, the sensor reads closer to room temperature, causing an error.

Is temperature calibration covered by SINGLAS at TIS?

No. The TIS SINGLAS accreditation covers pressure. TIS calibrates temperature instruments with traceable reference standards and issues a traceable certificate.