By TIS Instruments engineering team  ·  27 November 2026  ·  6 min read

It is 7 am, the plant is about to start up, and the control room reports that a pressure reading "looks wrong". The transmitter has been in service for a year, nobody remembers its last calibration, and the operators want an answer before the shift starts. Situations like this are exactly why a clear, repeatable transmitter calibration routine matters.

To calibrate a 4–20 mA pressure transmitter, you apply known pressures across its range with a reference calibrator, measure the output current at each point, and compare it with the ideal value – 4 mA at the lower range value and 20 mA at the upper range value. If the error is outside tolerance, you adjust zero and span and test again. In this guide, our engineers walk through the full process, with a worked example and the mistakes we see most often in the field.

Key takeaways

  • A transmitter is checked by applying known pressures and reading its mA output, not just by looking at the DCS value.
  • Ideal output = 4 + 16 × (pressure − LRV) ÷ (URV − LRV).
  • Test at 0, 25, 50, 75 and 100% of range, going up and coming down.
  • Express the error in % of span and compare it with the transmitter's tolerance.
  • Record "as found" before adjusting, and "as left" after.

Why Pressure Transmitters Need Regular Calibration

A pressure transmitter is the eyes of the control system. Its 4–20 mA signal drives alarms, control loops and sometimes safety trips. Over time, every transmitter drifts. Temperature cycling, vibration, pressure spikes and simple ageing of the sensor all move the output a little away from the truth.

The problem is that drift is silent. The signal still looks reasonable, so nobody questions it until a batch goes off-spec, a pump runs dry or an auditor asks for records. Regular calibration catches drift early and gives you evidence that the measurement can be trusted.

What You Need Before You Start

Good results start with the right equipment. For most process transmitters you need:

  • A pressure source – a hand pump for occasional work, or an electric or automatic pressure calibrator if you calibrate many transmitters.
  • A reference – a calibrator or test gauge at least four times more accurate than the transmitter. (Our guide to pressure accuracy and the 4:1 rule explains why.)
  • mA measurement and 24 V loop power – many TIS calibrators measure 0–24 mA at 0.01% FS and supply loop power, so no extra meter is needed.
  • Hoses and fittings for the transmitter's process connection.
  • A HART communicator if you need to trim a smart transmitter digitally.
TIS APC-300 automatic pneumatic pressure calibrator with mA input for transmitters
An automatic calibrator such as the TIS APC-300 generates pressure, holds each point and reads the transmitter's mA output in one set-up.

Understanding the 4–20 mA Signal

A transmitter maps its calibrated range onto a current between 4 and 20 mA. The 4 mA "live zero" lets the control system tell the difference between a real zero reading and a broken wire (0 mA), and leaves enough current to power a two-wire transmitter.

For a transmitter ranged 0–10 bar, the ideal outputs are:

Applied pressure % of span Ideal output
0 bar 0% 4.00 mA
2.5 bar 25% 8.00 mA
5 bar 50% 12.00 mA
7.5 bar 75% 16.00 mA
10 bar 100% 20.00 mA

The general formula is ideal mA = 4 + 16 × (P − LRV) ÷ (URV − LRV), where LRV and URV are the lower and upper range values.

Step-by-Step Calibration Procedure

1. Prepare and isolate safely

Get the permit, inform the control room, and put the loop in manual or bypass so the calibration does not upset the process. Close the isolation valve and vent the transmitter.

2. Connect the equipment

Connect the pressure source and reference to the transmitter's pressure port. Connect the mA input in series with the loop, or use the calibrator's own 24 V loop power with the field wiring disconnected.

3. Warm up and leak check

Allow the transmitter and calibrator their warm-up time. Then raise the pressure to about 100% of range and hold it for a minute. The reading should stay steady; if it falls, find the leak before going further.

4. Exercise the transmitter

Take it to full scale two or three times and back to zero. This settles the sensing element and gives more repeatable results.

5. Record the "as found" readings

Apply 0, 25, 50, 75 and 100% going up, then the same points going down. At each point, wait for the pressure to settle and record the reference pressure and the mA output.

Pro tip

Approach each test point from the same direction and let it settle for a few seconds. Overshooting and coming back introduces hysteresis into your readings.

6. Calculate the error

Express each error as a percentage of span: error % = (measured mA − ideal mA) ÷ 16 × 100. Compare it with the transmitter's specified accuracy or your site tolerance.

7. Adjust zero and span if needed

If any point is out of tolerance, trim the zero at 0% and the span at 100%, either with the transmitter's adjustment or with a HART communicator. Then repeat the five-point test.

8. Record the "as left" result

Document the final readings, the reference serial number and its certificate number, the date and the technician. Return the transmitter to service and tell the control room.

Worked Example

A 0–10 bar transmitter has a tolerance of ±0.25% of span. At 50% (5 bar), the ideal output is 12.00 mA and the transmitter reads 12.03 mA.

Error = (12.03 − 12.00) ÷ 16 × 100 = +0.19% of span – inside the ±0.25% tolerance, so this point passes. If the 100% point read 20.06 mA (+0.38%), you would adjust the span and test again.

TIS EPC300 portable electric pump pressure calibrator
For field transmitters, a portable electric pump calibrator such as the TIS EPC300 reads mA with switchable 24 V loop power.

Common Mistakes We See in the Field

  • Re-ranging instead of calibrating. Changing the range in a HART communicator does not correct the sensor. A real calibration needs a known pressure.
  • Head errors. With liquid in the lines, a height difference between the reference and the transmitter adds pressure – about 0.098 bar per metre of water.
  • Ignoring the low side on DP transmitters. Vent the low side to atmosphere (or apply the correct pressure) as your procedure says.
  • Blaming the transmitter for a loop problem. Check wiring, power supply and the DCS scaling too.

Good to know

If you calibrate many transmitters every week, automation pays back quickly. An automatic calibrator reaches and holds each point, reads the mA output and stores the results, so technicians spend their time on adjustments, not on pumping. Read more in Automatic vs Manual Pressure Calibrators.

Conclusion

Transmitter calibration is not complicated, but it rewards discipline: the right reference, a leak-free set-up, five points up and down, a clear error calculation and honest records. Do that consistently and the "is this reading right?" conversation at 7 am becomes a quick check instead of a guess.

Frequently Asked Questions

How often should pressure transmitters be calibrated?

Most transmitters are calibrated every 12 months. Critical, safety-related or high-drift loops may need 6 months. See how often to calibrate pressure instruments.

What is the difference between a sensor trim and a range change?

A range change decides which pressures map to 4 and 20 mA. A sensor trim corrects the transmitter's measurement against a reference. Calibration needs a trim against a known pressure, not just a range change.

Can I calibrate a transmitter without removing it?

Often yes. With isolation valves and a test connection, you can calibrate in place using a portable calibrator with loop power. Difficult installations are sometimes removed to a workshop bench.

Which TIS calibrators read 4–20 mA?

TIS automatic, electric pump and handheld PPC calibrators read mA signals and supply 24 V loop power, so you can apply pressure and read the transmitter in one set-up.